Gas chromatography columns
Packed columns made of glass or steel contain a stationary phase , either loaded directly into the column if it is a solid at its operating temperature, or coated onto the surface of a solid support if it is liquid at its operating temperature. Thus, the operating principle of GC can be distinguished as either gas–solid chromatography (mainly an adsorptive process by the stationary phase) or gas–liquid chromatography (GLC; mainly partition of the analytes between the mobile and stationary phases), based on the physical characteristics of the stationary phase. Capillary columns, introduced in the early 1980s, have now replaced packed columns for most applications. The original glass columns were fragile and have been superseded by fused-silica capillary columns. Fused silica is high-purity synthetic quartz, with a protective coating of polyimide applied to the outer surface. Since these columns retain their flexibility only as long as the coating remains undamaged, their operating temperature must be maintained below 360C for standard columns (400C for high-temperature polyimide coatings). The first capillary columns were 0.2, 0.25- or 0.32-mm internal diameter (i.d.) and between 10 and 50 m long. Subsequently ‘mega bore’ (0.53 mm i.d.), ‘minibore’ (0.18 mm i.d.) and ‘microbore’ (0.1 or 0.05 mm i.d.) columns have been developed. When coated with a heat-resistant polymer, these have the advantages of flexibility and strength, and can be threaded with ease through intricate pipework. A single column can be fitted into almost any manufacturer’s GC apparatus. Capillary columns provide improved resolution, sensitivity and durability, less bleed with increasing temperatures, and ease of maintenance and repair; they yield reliable and highly reproducible separations, typically over many hundreds or thousands of injections. This has reduced the number of columns required to achieve a satisfactory separation of quite complex mixtures.
The internal surface of the silica is deactivated by a variety of processes that can react silanol groups (Si-OH) on the silica surface with a silane reagent (usually a methyl or phenyl methyl surface is created). For gas–solid capillary chromatography, a fine layer (usually less than 10 μm) of stationary phase particles is adhered to the tubing (porous layer open tubular columns, or PLOT columns). For gas–liquid capillary chromatography, the stationary phase may be coated or covalently bonded directly onto the walls of the column (wall-coated open tubular columns, or WCOT columns) or onto a support (e.g. microcrystals of sodium or barium chloride) bonded to the column wall (support-coated open tubular columns, or SCOT columns). Stainless-steel capillaries are usually reserved for applications that require extremes of tempera ture, or where the possibility of column breakage cannot be tolerated. Nowadays, the internal surface of metal columns is specially deactivated chemically or by lining with fused silica, which allows for more flexibility and greater durability and reduces the possibility of degradation of analytes which can be catalysed by the presence of metal surfaces. Both bonding and cross-linking impart enhanced thermal and solvent stability to the stationary phase (often designated ‘DB’), and should be used if they are available. For conventional packed column chromatography, however, the stationary phase must first be coated onto the surface of a solid support.